Photochemical reaction apparatusCT-GUX60security:
The following basic requirements must be observed in all phases of operation, maintenance and instrumentationsecurityMeasuresu3002 When the instrument is used, it should be operated according to the instructions,Illegal use will result in the normal operation of the instrument and damage to the instrument.
AIn order to avoid electric shock accidents, input power line equipment must be grounded, the use of the instrument is three core grounding plug, the plug is connected with the anchor, if the plug can not be inserted into the seat, it should be an electrician to install the correct socket, don't make the instrument lose ground protection function.
BPay attention to using the power supply:
Before connecting the AC power supply, it is necessary to ensure that the voltage is consistent with the required voltage of the instrument±10%Deviation and ensure that the rated load of the power outlet is not less than the instrument requirements.
CPay attention to using the power supplyLine:
The instrument usually uses a random power supplyLine. If the power supplyLineDamaged, must be replaced, not repaired. The same type and size of power supply must be used in the replacementLineSubstitute for. When the instrument is used, the power supplyLineDon't put anything on it. Don't take the powerLineIn places where people walk around.
DPay attention to instrument placement:
The instrument should be placed in a cool, ventilated, dry, dustproof good position, in order to better heat dissipation effect, the instrument ventilation, other items should keep the effective distanceN﹥30cm).
Photochemical reaction apparatusCT-GUX60Application range:
Photochemical reaction apparatusCT-GHX60A new generation of photochemical reaction device has been developed with the cooperation of famous laboratories in china,Photochemical reaction apparatusCT-GHX60It is mainly applicable to the study of gas phase, liquid phase, solid phase and flow system under UV irradiation, simulated sunlight, visible light, and special analog light irradiationTiO2Photochemical reactions, such as photocatalysts, for chemical synthesis, environmental protection and life sciences. The system has the advantages of reasonable technology, simple structure, touch button, convenient operation, stable operation, protection of the human body, free combination, flexible customization and other unique advantages.
Photochemical reaction apparatusCT-GUX60Product features:
1.Photochemical reaction apparatusCT-GHX60The main control power supply, flexible control of lighting time, suitable for timing operation and experimental use;
2.Professional stable analog light source and stable, space saving volume design, especially suitable for space limited laboratory equipment;
3.There are many test tube magnetic stirring reactor functions in the supporting device, which can realize some test tube aeration function at the same time. The multi test tube magnetic stirring reactor has excellent performance value in practical application;
4.There are many supporting magnetic stirring reactor, can be a strong magnetic stirring, inflation, deflation, sealed in the reaction process, the function of measuring
5.The solid reaction device can be used to catalyze the solid substance and enhance the catalytic speed;
6.Equipped with cooling water device, imported compressorFluorine free operation, ensure the light for a long time and stable operation, suitable for continuous operation test, the cooling water supply device provided with a mute external circulation pump, cooling water circulation pressurization, while saving water waste;
7.Photochemical reaction apparatusCT-GHX60The special agitator is a new special strong magnetic stirring device, which can make the sample fully and evenly absorbed by light, and realize simultaneous stirring of multiple test tubes at the same time. The stirring magnet can be stepless speed regulating in each test tube.
8.The main components of excellent selection, selection of DC motor of high power motor, stirring force is huge, small noise, the strongest magnetic magnet "neodymium, iron, and" water magnetic rotor, ensure sufficient torque and sharp,
Photochemical reaction apparatusCT-GUX60 Advantages and characteristics
1.There are many light source devices, glass reaction containers, all quartz light source protection cold trap, dedicated parallel.The rotary reaction instrument, the main reaction protection box and the volume control device are made up of parts.
2.It can be used for the determination of reaction kinetics and reaction constants, the analysis of reaction products and quantum yield, and the development of green energy.
3.The whole quartz reaction vessel is divided into two parts:A,Btype,Meeting different customer requirements,(otherwise accepted by non-standard reactors)
4.The high and low temperature circulator provides a light source to protect the cold trap from cooling or constant temperature reaction,At the same time, according to the different power light source flexible collocation, improve the most efficient high low temperature constant temperature circulation device,(-20degree-100degree)
5.The ultraviolet filter device is equipped to ensure the optimization of light source.
6.booktypeNumber configuration can recommend the optimal combination configuration according to the actual use of customers,It can also be tailored for customers to meet the actual requirements of the photochemical reaction device
Photochemical reaction apparatusCT-GUX60 Configuration list
One,SpecificationstypeNumber:CT-GHX60
Two,Product name:Photochemical reaction apparatus
Three,Configuration standard:1.The main part of the reaction includes the camera obscura,Mercury lamp electric controller,Xenon lamp electric controller,Glass reactor,light source,A magnetic stirreruff1b
2.Mercury lamp electric controllerIn order to control300W,500WWorking conditions of medium pressure mercury lamp;(Optional100W,200W,1000W,1500Wetc.Specifications)
3.Xenon lamp electric controllerIn order to control500W,1000WWorking conditions of xenon lamp;(Optional250W,350W,800W,1500Wetc.Specifications)
4.All quartz glass reactors can be used separately20ML,50ML,60ML,100ML,300ML,500ML,1000MLetc.
5.All silica glass cold trap(Cooling bottom extends the cooling cycle)
6.To configurePhotochemical reaction apparatusCT-GHX60Dedicated parallel reactionA magnetic stirreruff1b
7.The cooling water circulation device can be realized-20degree-100degreePrecise temperature control.
Photoelectric parameters
High pressure mercury lampPhotoelectric parametersas follows:
powerW
|
Starting currentA
|
Working currentA
|
working voltagev
|
Effective arc lengthMM
|
meetLinemode
|
100
|
1.5
|
0.4
|
145.8
|
75±10
|
Single end extraction
|
300
|
5.7
|
3.4
|
138.1
|
125±10
|
Single end extraction
|
500
|
7.1
|
5.4
|
140
|
170±10
|
Single end extraction
|
1000
|
17.3
|
10.4
|
143.2
|
250±10
|
Single end extraction
|
High pressure mercury lampRelatively strongdegreeas follows:
wavelengthnm
|
250
|
313
|
365
|
400
|
510
|
620
|
720
|
Right now strongdegree%
|
20
|
85
|
100
|
30
|
20
|
40
|
80
|
Xenon lampPhotoelectric parametersas follows:
Xenon lampabroadtypeSize andPhotoelectric parameters
powerW
|
Starting currentA
|
Working currentA
|
working voltageV
|
Effective arc lengthMM
|
meetLinemode
|
350
|
5.5
|
5.0
|
73.1
|
190±10
|
singleLinemode
|
500
|
8.3
|
6.4
|
75.6
|
190±10
|
singleLinemode
|
800
|
11.4
|
10.0
|
83.5
|
220±10
|
singleLinemode
|
1000
|
12.4
|
12.1
|
87.6
|
220±10
|
singleLinemode
|
mercuryXenon lampPhotoelectric parametersas follows:
mercuryXenon lampabroadtypeSize andPhotoelectric parameters
powerW
|
Starting currentA
|
Working currentA
|
working voltageV
|
Effective arc lengthMM
|
meetLinemode
|
300
|
6.3
|
4.1
|
220±20
|
175±10
|
singleLinemode
|
500
|
9.7
|
6.2
|
220±20
|
190±10
|
singleLinemode
|
1000
|
19.1
|
11.7
|
220±20
|
200±10
|
singleLinemode
|
Metal halide lampPhotoelectric parametersas follows:
Metal halide lampabroadtypeSize andPhotoelectric parameters
powerW
|
Starting currentA
|
Working currentA
|
working voltageV
|
Effective arc lengthMM
|
meetLinemode
|
250
|
4.8
|
3.0
|
80.7
|
35±10
|
singleLinemode
|
400
|
10.7
|
5.4
|
80.2
|
35±10
|
singleLinemode
|
800
|
26.73
|
10.3
|
90.9
|
45±10
|
singleLinemode
|
1.mercuryLamp, xenon lamp open5After a few minutes, the light is stable and the lights need to be turned offetc.stay5Restart in minutes;
2.mercury,Xenon lampThe luminous part is quartz tube,Hand straightmeetmeetContact will cause pollution and affect the luminous intensitydegree,quartz
WareabroadClean with a small amount of alcohol,Wash the interior with lotion;
3.Selection of reaction tubes for multi tube magnetic stirrer30ml,50ml,60ml,100ml,300ml,500ml,1000mletc..
4.Main box,Main boxThe place where a series of chemical reactors are placed,With the following characteristics:Box
Size is440mm*440mm*770nmabout26KG
5.Ambient temperaturedegree:0℃-35℃,Relative humiditydegree≤80%.
6.Defects Liability Period:Oneyear