1 PA30PA30A digital leakage current tester overview
This series of equipment is a true RMS digital intelligent instrument for measuring the leakage current between the power supply end of any applicable to the measurement of various single-phase motors, electrical appliances, instruments, household appliances and other polar and protective earthing terminal. The instrument has the pre setting function, convenient use, flexible operation, and according to user request communication interface. Is in line with the GB4706.1-2005 "Test safety of household and similar electrical appliances - General requirements" and other relevant national standards in terms of the requirements of the test equipment needed. This instrument is carried out Q/TPGG3PA30 Leakage current tester series digital enterprise standard.
2 PA30PA30A digital Leakage Current Tester Specification and technical characteristics
See Table 2.1 specifications and main technical parameters One .
Really effective
The basic value of measurement error. (2%r+2d)
2.4 product features
A true RMS measurement, automatic range conversion.
The B panel type correction software.
C memory test parameters.
D single or continuous test, suitable for the production line.
E has the over-voltage and over-current protection function.
F control interface cable. Optional communication interface.
A avoid placement of direct sunlight, rain and moisture.
B please away from fire and heat, to prevent the high temperature of the machine.
C handling or maintenance, should be shut down and will remove the power line.
5 use and operation
5.1 of the initial state
Connect the instrument power supply , Open the power switch , The test voltage for display" Zero " V "; leakage current" display for" Zero " MA Display for time " 00 " S ; automatic or manual indicator in the last set, the instrument is in the initial state.
5.2 function set
5.2.1 test voltage setting
APA30
Press the "measure" button, turn the voltage adjust knob, the voltage display indicates the measured object is the rated voltage of the One point zero six Times or One point one (e.g. 233V or 242V ), this voltage test voltage and keep the voltage adjust knob position unchanged. After the set, press the "reset" button.
BPA30A
Users in the" N "And" L Direct access to "end with the power supply voltage test.
5.2.2 each phase measurement time setting
In the initial state, press "set" button, "test voltage" for display" TEST "," time "display the last set value, enter each phase measurement time setting. According to the "long" increasing key per phase measuring time value, in descending bond amounting to reduce the measuring time value of each phase. Setting range 1s~99s (including 0s Said measuring time is infinite, i.e. manual commutation measurement). Press "set" button to save each phase measurement time setting and into the next set state again ( The current alarm set ) . If press the "reset" button to cancel the next set, exit the setting, the instrument for the initial state.
5.2.3 manual / Automatic commutation mode selection
If every time is not set for phase measurement" Zero "When is the" automatic "commutation mode," automatic "indicator light, when to press the" measure "button" N Phase measurement, phase measurement after each time, automatically" L Phase measurement; if each phase measurement time is set to" Zero "When is the" manual "commutation mode," manual "indicator light, when press the" measure "button, can press the" phase change "button to change the phase measurement.
The current alarm set 5.2.4
In the initial state two press "set" button, "test voltage" display for" ALAr "," leakage "display the last set value, enter the current alarm setting state. According to the "long" increasing bond increased the current alarm set value, in descending
6 bond amounting to reduce the current alarm set value. Setting range 0~20.00mA . Press "set" button to save the current alarm set value and into the next set state again ( Address code setting ) . If press the "reset" button to cancel the next set, exit the setting, the instrument for the initial state.
5.2.5 address code setting (only for communication settings)
In the initial state, according to the three "set" button, "test voltage" for display" Addr "," time "to display the last address code setting, namely equipment into the address code setting state key" / "increased according to increasing address code value, in descending bond amounting to reduce the address code value, setting range 0~99 Again, press the "set" button, that holds the address code setting and into the next set of state ( Baud rate setting ) At this time, if you press the "reset" button to cancel the next set, exit the setting, the instrument for the initial state.
5.2.6 baud (only for communication settings)
In the initial state, according to the four "set" button, "test voltage" for display" BPS "," leakage current "display the last baud rate setting value, enter the baud rate setting state. According to the "long" to increase the baud rate increasing key value in decreasing bond amounting to reduce the baud rate, the set value is" 1200/2400/4800/9600 "Four files, press" set "button to save the baud rate setting value and complete set again, exit the setting, the instrument for the initial state.
5.3 testing steps
5.3.1 complete set, the instrument is in the initial state.
Seven
If the a object is a single-phase three plugs are not connected "measuring end I "Only the measured object is directly inserted into the instrument on the back plate
The measuring end of the socket.
If the B object is two plugs, in addition to the measured object is directly inserted into the instrument on the back plate of "measuring end II", and use the test line
The measured object protection grounding terminal and instrument on the back plate of "measuring end of" properly connected.
If the C does not use isolation transformer, the measured object must be insulated and open the earth (only for PA30A ).
Note: the "end of measurement" and "measuring end II" has been connected in the instrument.
5.3.3 status indication
When measuring, press the "measure" button, the instrument began tested, this "test N "Light instrument first on the measured object" N "Test" at the same time, the test voltage display shows the current voltage test, leakage current display leakage current, "" time "to display the remaining test time, such as the measured value of leakage current is greater than the set value, the alarm lights, after the test ended, the buzzer sends out alarm sound. At the same time" N "The lights, alarm lamp and the current value of the latch, no longer measuring instrument" L On the other hand, such as "phase; the measured leakage current is less than the set value, the test time after the end of" qualified "lights and the current value of the latch," N "After testing, automatic measuring instrument of commutation" L "(the content is based on the" automatic "indicator light in the case; such as the" manual "indicator light, need to finish" N "After, press the" phase change "button to start measuring" L "). When the two time after measurement showed the maximum value, such as qualified, instrument issued a short "beep" sound.
5.3.4 protection function
When the test voltage exceeds the 300.0V The display will show "voltage" test" ---- "Said overvoltage. The leakage current of more than 21.00mA The leakage current, the display will show "" ---- "Said the current. You should stop testing and instrument reset". When the test voltage exceeds 8" Three hundred and thirty " V The leakage current exceeds or" Twenty-two " MA Stop, automatic measuring instrument, enter the state of protection, the display flashes, waiting for the "reset".
6 communication protocol
6.1 baud rate is set by a panel key" 1200/2400/4800/9600 " BPS Four
6.2RS485 Pinout: the standard level RS485 level
Pin definition:
Eighty-nine
AB
6.3 frame format: One An initiation +8 Data bits +1 Stop bit
6.4 code sequence: address code + Instruction code + Data code
Note: data code ASCII code
6.5 communication protocol
A pc:
(1) by PC RS485 Export to the instrument issued a string of code byte
The first byte: address code" #01H "As the instrument address code
Second bytes: instruction code" #10H "As the" start "signal code
" #11H "As a" reset "signal code
" #20H "As the" parameter setting "code
" #21H "As the request of the loopback test data and the results of instruction instrument
(2) third ~ eleventh bytes for the parameter setting data in high byte code, low byte in third, fourth bytes:
" TIMESET " ASCII Data code, such as" #39H "," #31H ( Ninety-one )
Fifth, sixth bytes:" CHANGTIMESET " ASCII Data code, such as" #30H "," #35H (" 05 )
Seventh to the cross section:" ALARMSET " ASCII Data code, such as" #31H "," #39H "," #38H "," #30H "
( One thousand nine hundred and eighty )
11 byte:" ALARMSET " ASCII The decimal point data code
" #30H "As a decimal point in
" #31H "In the ten decimal point position
" #32H "As the decimal point in 100 on
" #33H "As the decimal point in a thousand
Such as: Nineteen point eight zero
B instrument
When the instrument receives the PC from the corresponding address code, the corresponding operation according to the corresponding instruction code and data codes, when the instruction code for the requested data and results (loopback test instrument" #21H "), to the PC continuous echo measurements and results ( ASCII The high byte code): first, low byte in the post.
Order code: address code + Data code
Nine
The first byte: address code" #01H "(instrument address)
Second ~ fifth bytes: N By test data, such as" #30H "," #31H "," #38H "," #36H (" 0186 )
Sixth bytes: N By test data of the decimal point:
" #30H "As a decimal point
" #31H "As the ten decimal place
" #32H "As the decimal point 100
" #33H "As the decimal 1000
Seventh bytes: N Is the test data to determine
ASCII code" P Qualified ":
ASCII code" F "Not qualified
ASCII code" N Results: no initial state
Eighth ~ eleventh bytes: L By test data, defined as second to fifth bytes
Twelfth bytes: L The test data is the decimal point, defined as sixth bytes
Thirteenth bytes: L Phase testing results of judgment, defined as seventh bytes
7 wire interface pin definition:
notes One "Qualified", "alarm" output contact signal (control), effective closure.
notes Two Output: contact can bear 24V/0.2A The power supply.
notes Three "Qualified", "alarm" signal output contacts are "automatic commutation mode and test time to
Rear action.
notes Four "Start" and "reset" as the input signal contacts (controlled). among Two As a common foot.
8 common faults and Solutions
The 8.1 is powered up, the power indicator light and display are not bright.
Check the power fuse on the rear panel is damaged, such as fuse fuse replaced by the same type.
8.2 press the "measure" or "phase change" button "test N "Or" test L "Lights, adjust the voltage adjust knob,
"Voltage" display voltage value is displayed, but the measuring terminal output voltage.
Voltage transformer is damaged, replace the same types of components.
8.3 press the "measure" or "phase change" button "test N "Or" test L "The lights are lit, adjusting the voltage adjust knob,
"Voltage" display voltage values are displayed at the same time measuring end no voltage output.
Start pass alarm
One hundred and twenty-three million four hundred and fifty-six thousand seven hundred and eighty-nine
reset
Contactor 10 within the instrument is damaged, replacement of the same model element.
8.4 "no current value measurement, current display.
A test line open circuit or measuring the connecting end on the plug socket contact, check the parts above, replace the same damage element.
B resistance in the circuit or special module is damaged, replacement of the same model element.
8.5 other issues: please contact directly with the company.
9 test methods
9.1 test voltage test
Instrument voltage value ( V X ) and standard electronic voltmeter output voltage (V ), the relative error calculated should be consistent with the basic error voltage test in table 1.
Eleven
9.2 leakage current test
chart Eight Leakage current connection
According to the chart Eight The wiring, instrument to measure state respectively. N "Or" L ", a rising voltage regulating resistor slowly increases the current, the current value of observation instrument (IPA30PA30A Digital leakage current tester ) with standard digital ammeter display on the corresponding value (I ), the relative error calculated in accordance with table One The leakage error of PA30PA30A type digital current leakage current tester
10 complete
PA30A or PA30 Digital leakage current tester of a PA30PA30A type digital leakage current tester
A wire plug
Can not reconnect the power line (5kVA The instrument is not attached ). A
A test line
Three core power plug ( only PA30-3kVA , 5kVA attach ) A
Use a manual
Product certificate
11.1 the instrument should be to keep the temperature in the room Zero C ~ Forty C, the relative temperature not exceeding 80% In the air, and not enough to contain harmful substances caused by corrosion.
11.2 instruments from the user purchase date Twelve Within a month, when users use fully comply with the rules in the specification, and the user does not attempt to change the instrument internal structure, it is found that the instrument is not working properly, then the manufacturer should be responsible for free to replace or repair.
11.3 manufacturing plants have the right to make changes without notice to the product.
Thirteen
L EFUP
Poisonous and harmful substances or elements contained in the instrument is not used in the normal conditions, leakage or mutation, users use the instrument will not cause serious pollution to the environment or causing serious damage to the person and property of the period Ten Year. The poisonous and harmful substances or elements of identity name and content table.
Schedule of poisonous and harmful substances or elements of the name and content identification
Part name
The poisonous and harmful substances or elements
A body of XXX, XXX
Shell plating XXX X Eighty
LCD screen X Eighty X Eighty
Circuit board X XXX, eighty
Contactor X XXX, eighty
Transformer
(pressure regulator )
X XXX eighty
switch X XXX, eighty
Ten: the concentration of the hazardous substance in all homogeneous materials in the parts are in SJ/T11363-2006 The standard limits for the following.
X said: the limited requirements of the content of toxic and harmful substances in at least one of the homogeneous materials beyond the SJ/T11363-2006 standard.
Note: This product 80% Manufacture of components using non-toxic environmentally friendly materials, instead of containing toxic and hazardous substances or elements of parts because of global technical level limit and cannot achieve the toxic and hazardous substances or elements.