LS126A? UV radiometer An instruction manual? V2

???? LS126A UV radiometer professional measurement of UVA radiation can power density, i.e. ultraviolet radiation power per unit area.
Suitable for measuring the light intensity of ultraviolet radiation curing process, guide the user to select the right power curing light, light curing light and judgment condition, the curing process was optimized, improve the curing effect.
The aging of lithography exposure, ultraviolet light source, integrated circuit lithography, ultraviolet radiation, photochemistry and polymer materials,
The UV irradiance measurement, medical testing, optical communication and other fields.
LS126A? The? The structure and parameters
1. spectral response: 320nm-400nm
2. range: 0-2000 mW/cm2
Measurement accuracy: + 3. 10%
Four Resolution: Zero point zero one MW/cm??
5. main dimensions: length? 130mm * W? 70mm *? 28mm
6. probe size: diameter? 45mm * thickness? 12.5mm
7. probe length: 1? rice
Instrument weight: about 8.? Two hundred and fifty? gram
Nine Battery: Four section AAA Alkaline dry battery.
?
Measurement and precautions
One??? The instrument probe receiving window is ultraviolet light, ultraviolet radiation can be obtained the test point to the power density,
with? When the maximum ( MAX The current value (and) RT ).
Two??? For the measurement of light curing, the probe back close to the conveyor belt, window to the receiving probe, probe follow belt together
After UV light irradiation area, power density can be obtained in UV lamps. The instrument probe (bring back magnet, can
The adsorption on the iron plate, convenient fixing)
Three? When not in use, please click"??? "The key to shutdown.
Four? Avoid contact with corrosive materials, away from the high humidity environment.
Five? Please put it off after the special packing, safekeeping.
Six? Check: check the cycle cycle is recommended for one year.