Product function:
K, Na can be tested at the same time, can also test lithium.
LCD display, digital direct reading.
The size of the flame has preselected function and automatic flameout protection device.
It has the advantages of simple operation, fast analysis, less samples.
- use liquefied gas as fuel.
The main technical indicators:
The stability of continuous injection with standard solution, 15s instrument shows the value of relative maximum variation is less than or equal to 3% were measured 1 times per minute, determination of the relative maximum variation is less than or equal to 15% times the instrument shows the value of 6
The repeatability: repeated 7 times less than 3% independent measurements on the same standard solution
The linear error: K = 0.005mmol/L: (0.0100 ~ 0.0800) mmol/L: Na = 0.03mmol/L (0.0500 ~ 0.400) mmol/L
The detection limit: K: Na: detection limit is less than or equal to 0.004mmol/L; the detection limit is 0.008mmol/L
The response time: the response time is less than 8S
The sample aspiration volume: sample aspiration volume less than 6mL/min
Other:
- Complete: a mainframe; a compressed air compressor; preparation of a set of accessories
The normal working conditions: Ambient temperature: 10 to 35 DEG C Relative humidity: less than 85% The product should be placed in the table without vibration, avoid direct exposure to bright light, no strong electromagnetic interference,
Influence of strong airflow, without affecting the use of vibration. The product use site should not be flammable and explosive, corrosive gases and equipped with fire fighting equipment. Supply voltage: 220 + 22V, frequency: 50HZ + 1 HZ, and
Have a good grounding Power rating: 250W