- 如果您对该产品感兴趣的话,可以
- 产品名称:PH/ ion concentration meter
- 产品型号:MP523-01
- 产品展商:Shanghai Sanxin instrumentation
- 产品文档:无相关文档
简单介绍
According to the international standard GLP requirements, it has automatic calibration, automatic temperature compensation, data storage, RS232 output, clock display, function settings and self diagno
产品描述
| In conformity with the international standard GLP requirements, with automatic calibration, automatic temperature compensation, data storage, RS232 output, clock display, function setting and self diagnosis information intelligent function. |
| We have three different measuring modes: stable display mode; timing measurement mode; continuous measurement mode. |
| The automatic identification of 13 kinds of pH standard buffer solution, a standard buffer solution four series can choose: Europe and the United States series, NIST series, Chinese series and custom solutions. |
| Can be set high pure water pH measurement mode and ammonia water pH measurement mode, electric power, petrochemical and other industries to meet the use requirements. |
| - EH redox potential measurement mode, display directly relative to the standard hydrogen electrode potential value of ORP. |
| 10 kinds of instruments - ion mode commonly used storage options, also can set their own other ions. |
| The four switchable ion concentration: pX, mol/L, mg/L unit and ppm. |
| The instrument with unique serial number. |
| The IP54 dust splash grade, socket with silicone cap seal protection. |
|
PH
|
measuring range
Resolving power
Accuracy
Input current
Input impedance
stability
Temperature compensation range
|
(-1.999 ~ 19.999) pH
0.1/0.01/0.001 pH
Electric meter: + 0.002 pH + 0.01 pH, supporting:
Less than 1 * 10-12 A
Over 3 * 1012.
+ 0.002 pH/3h
(0 ~ 100) C (automatic or manual)
|
|
MV
|
The measuring range (mV/ORP/EH)
Resolving power
Accuracy
|
-1999.9mV ~ 0 ~ 1999.9mV
0.1mV
+ 0.03% FS
|
|
Ion concentration
|
measuring range
Display unit
Accuracy
Temperature compensation range
|
0 ~ 14 pX
PX, mol/L, mg/L, PPM
Electric meter reading: + 0.5% (I), 1% readings (two price)
0 ~ 60 C (automatic or manual)
|
|
temperature
|
measuring range
Resolving power
Accuracy
|
-10 ~ 110 DEG C
0.1 C
5~ 60 DEG C: + 0.4 + - 0.8 DEG C rest range
|
|
Other technical parameters
|
Data storage
Storage content
?
Power Supply
communication interface
Size and weight
Quality and safety certification
|
The 900 group
Measurements of number, measurement, temperature, ATC or MTC status, measuring date and time measurement
DC9V/300mA
RS232
160 * 190 * 70mm/880g
ISO9001:2000, CE and CMC
|
|
Working conditions
|
ambient temperature
Ambient humidity
IP grade
|
5 ~ 35 C (0.01) 15 ~ 30 C (0.001)
Less than or equal to 75%
IP54 dust and splash
|