Host function:
The measurement system based on microcomputer, T-A Conversion accuracy, automatic adjustment 0% (T) and 100% (T), with GOTO lambda, automatic 8 Combined sample holder, automatic deduction error cuvette, concentration of multi point calibration, slope and intercept settings and other functions, has RS232 Interface and parallel printer port (printer, separately optional).
Online function:
adopt RS232 The serial interface, can be used for full wavelength scanning, using the software time to scan, concentration (regression equation 1 times, 2 times, 3 times ), multi wavelength test, and greatly expand the test function (except the main function).
Technical indicators:
The wavelength range: 200nm ~ 1000nm
- source: tungsten halogen lamp 12V20W DD2.5 (import) Deuterium lamp (import)
The light receiving element: Battery
The maximum allowable error - wavelength : 0.5nm.
- wavelength repeatability : = 0.2nm
The spectral bandwidth: 2nm
The stray light That is less than or equal to 0.3% (T) (at 220nm and 360nm, to NaNO Two Determination)
The transmittance measurement range: 0% (T) to 200% (T)
The absorbance measurement range: - 0.301 (A) to 4 (A)
The transmission ratio of the maximum permissible error : + 0.5% (T)
The proportion of complex transmission That is less than or equal to 0.2% (T)
- noise: 100% noise is less than or equal to 0.5% (T) 0% noise is less than or equal to 0.2% (T)
- stability : = 0.004A/30min
Other:
- power: AC220V + 22V 50Hz + 1Hz