Shanghai Chi Tang Electronics Co., Ltd. 主营:
咨询电话:13661497600
产品详情

Gas chromatograph

  • 如果您对该产品感兴趣的话,可以
  • 产品名称:Gas chromatograph
  • 产品型号:GC126
  • 产品展商:Shanghai instrument electricity
  • 产品文档:无相关文档
简单介绍

A new design of GC126 microcomputer temperature control system, using the international advanced control technology. High precision (less than 0.1 DEG C) the reliability and anti-jamming performance.

产品描述

The characteristics and performance:
The new design GC126 The microcomputer temperature control system, using the international advanced control technology. High precision (better than + 0.1 C The reliability and anti-jamming performance).
The keyboard control and the use of various parameters, the machine has a self diagnosis, power-off protection, file storage and call, limit temperature setting function. 3.8" The big screen can display the various temperature set value and the actual value and the real-time curve, the detector parameters, signal value and real time curve.
Having six independent control areas: column box, injector, detector, auxiliary device.
The product can achieve nine - step temperature programmed, with automatic cooling function of pillar box door, achieve rapid heating and rapid cooling.
The instrument is equipped with double and single type high sensitivity hydrogen flame ionization detector (FID), One of the following optional detector, gas sampling valve, reformer and other accessories can be installed at the same time, two kinds of detector, and the detector of high sensitivity (e.g. FID The test results:??

D FID = 3 x 10 - Twelve G/s), stability in short time, the nozzle cleaning and convenient installation.

Technical characteristics:
- Embedded Thirty-two A multi processor and CPU System development, IT Interface technology, temperature automatic parameter setting, detector selection, detector operating parameters and the data of auto zero function, sensitivity and polarity of the digital set.
- Chinese menu operation interface, and can display multiple sets of parameters, the operation is very simple and convenient.
The computer control function, through all internal computer operation (through the instrument function USB Interface).
No need of external workstation and provides an analog output interface.

? Column box:
The temperature range of room temperature: 1 5 to 399 DEG C
Control precision C: + 0.1
The temperature programmed: 9 rank
The maximum heating rate: 60 C /min

Injector:
The maximum loading injector (three The 2 column sample, 1 A capillary column sample) or Two A capillary column injector.
The sampling unit: independent split / splitless injector system and column injector system

Detector:
- carry a maximum of three detectors
The hydrogen flame ionization detector (FID):
The temperature range of room 30 vinga 399 DEG C. The limit of detection; = 4 x 10 - Twelve G/s sample: C Sixteen

The micro thermal conductivity detector (-TCD):
The highest temperature of 300 C Sensitivity; More than 5000mv - ml/mg, C Sixteen
The electron capture detector (ECD):
The highest temperature of 300 C Radioactive sources: Sixty-three Ni10~12mci The minimum detection amount: Eight * 10 - Fourteen Sample: r666 g/ml;
- Flame photometric detector (FPD):
The temperature range of room temperature heating 7 to 400 DEG C The limit of detection; = 1 x 10 -11 G/s (P) = 1 x 10 -11 G/s (S)
Formaldehyde parathion

产品留言
标题
联系人
联系电话
内容
验证码
点击换一张
注:1.可以使用快捷键Alt+S或Ctrl+Enter发送信息!
2.如有必要,请您留下您的详细联系方式!
Copyright@ 2003-2025  Shanghai Chi Tang Electronics Co., Ltd.版权所有      电话:13661497600 传真: 地址: 邮编: