- 如果您对该产品感兴趣的话,可以
- 产品名称:MP525 type pH/ dissolved oxygen measuring instrument
- 产品型号:MP525
- 产品展商:Shanghai Sanxin instrumentation
- 产品文档:无相关文档
简单介绍
The laboratory for high precision measurement of pH and dissolved oxygen. It is suitable for biotechnology, aquaculture, power plants and environmental protection industries, and has wider application
产品描述
In conformity with the international standard GLP requirements, with automatic calibration, automatic temperature compensation, data storage, RS232 output, clock display, function setting and self diagnosis information intelligent function. |
We have three different measuring modes: stable display mode; timing measurement mode; continuous measurement mode. |
The automatic identification of 13 kinds of pH standard buffer solution, a standard buffer solution four series can choose: Europe and the United States series, NIST series, Chinese series and custom solutions. |
Can be set high pure water pH measurement mode and ammonia water pH measurement mode, electric power, petrochemical and other industries to meet the use requirements. |
- EH redox potential measurement mode, display directly relative to the standard hydrogen electrode potential value of ORP. |
The dissolved oxygen measurement mode with automatic temperature compensation, automatic compensation and automatic pressure salinity compensation function. |
Polarographic oxygen electrode of the new structure can simultaneously measure temperature, salinity and dissolved oxygen, the dissolved oxygen electrode with a special calibration set, the electrode polarization only 3~5 min. |
The instrument with unique serial number. |
The IP54 dust splash grade, socket with silicone cap seal protection. |
PH
|
measuring range
Resolving power
Accuracy
Input current
Input impedance
stability
Temperature compensation range
|
(-1.999 ~ 19.999) pH
0.1/0.01/0.001 pH
Electric meter: + 0.002 pH + 0.01 pH, supporting:
= 1 x 10 -12 A
= 3 x 10 Twelve Omega
+ 0.002 pH/3h
(0 ~ 100) C (automatic or manual)
|
MV
|
The measuring range (mV/ORP/EH)
Resolving power
Accuracy
|
-1999.9mV ~ 0 ~ 1999.9mV
0.1mV
+ 0.03% FS
|
dissolved oxygen
|
measuring range
Resolving power
Accuracy
response time
Residual current
Temperature compensation range
Salinity compensation range
Pressure compensation range
Automatic calibration
Electrode type
|
(0 ~ 20) mg/L (PPM) (0 ~ 200)%
0.1/0.01 mg/L (PPM) 1/0.1%
Electric meter: + 0.10 mg/L + 0.40 mg/L, supporting:
Less than 30 s (25 C, 90% response)
Less than 0.1 mg/L
(0 ~ 45) C (automatic)
(0 ~ 45) PPT (automatic)
(80 ~ 105) kPa (automatic)
Is water saturated air; air saturated water
Polarographic
|
temperature
|
measuring range
Resolving power
Accuracy
|
-10 ~ 110 DEG C
0.1 C
5~ 60 DEG C: + 0.4 + - 0.8 DEG C rest range
|
Other technical parameters
|
Data storage
Storage content
Power Supply
communication interface
Size and weight
Quality and safety certification
|
The 900 group
Measurements of number, measurement, temperature, ATC or MTC status, measuring date and time measurement
DC9V/300mA
RS232
160 * 190 * 70mm/880g
ISO9001:2008, CE and CMC
|
Working conditions
|
ambient temperature
Ambient humidity
IP grade
|
5 ~ 35 DEG C
Less than or equal to 85%
IP54 dust and splash
|