- 如果您对该产品感兴趣的话,可以
- 产品名称:MP526 conductivity / dissolved oxygen measuring instrument
- 产品型号: MP526
- 产品展商:Shanghai Sanxin instrumentation
- 产品文档:无相关文档
简单介绍
The laboratory for high precision measurement of conductivity and dissolved oxygen. Suitable for biotechnology, aquaculture, power plants and environmental protection industries.
产品描述
Conforming to international normsGLPRequirements include automatic calibration, automatic temperature compensation and data storage, RS232Intelligent functions such as output, clock display, function setting and self diagnostic information. |
There are three different measurement patterns:Steady display mode; timing measurement mode; continuous measurement mode. |
Automatic recognition8There are two series of standard solutions for the conductivity standard solution:Europe and America series and Chinese series. |
Advanced conductivity measurement techniques are used with constantK=1Conductive electrode Just a little calibration, you can meet0.5 S/cm mS/cmMeasurement accuracy requirements. ~ 200 |
- Switchable conductivity, TDSSalinity and resistivity.TDSAnd salinity are calculated by polynomial to ensure the conversion precision of full range. |
The model of conductivity measurement has the function of high pure water nonlinear temperature compensation. |
The measurement mode of dissolved oxygen has the function of automatic temperature compensation, automatic salinity compensation and automatic air pressure compensation. |
The polarographic dissolved oxygen electrode with a new structure can measure temperature, salinity and dissolved oxygen simultaneously. The electrode is calibrated with a special dissolved oxygen electrode, and the electrode polarization is only required3~5 min. |
The instrument has a unique product serial number. |
- accord withIP54Dust prevention and splash grade, instrument socket with silicone cap seal protection.
|
Conductivity
|
Measuring range
?
?
?
?
?
Resolving power
Accuracy
Temperature compensation range
Electrode constant
Reference temperature
|
Conductivity: (0.00~20.00) S/cm (20.0~200.0) S/cm
(200~2000) S/cm (2.00~20.00) mS/cm
(20.0~200.0) mS/cm (200~2000) mS/cm (K=10)
Resistivity: (0 ~ 100) M cm.
TDS: (0 ~ 100) g/L
Salinity: (0 ~ 100) ppt
0.01/0.1/1 S/cm 0.01/0.1/1 mS/cm
Electric meter: FSMatching: 0.80%FS + + 0.5%
(0 ~ 50) C (automatic)
0.1 / 0.5 / 1 / 5 / 10 / 50 / 100 cm-1
25 DEG, 20 DEG and18 DEG
|
Dissolved oxygen
|
Measuring range
Resolving power
Accuracy
Response time
Residual current
Temperature compensation range
SalinityCompensation range
PressureCompensation range
Automaticcalibration
Electrode type
|
(0 ~ 20) mg/L (PPM) (0 ~ 200)%
0.1/0.01 mg/L (PPM) 1/0.1%
Electric meter: mg/LMatching: mg/L + 0.40 + 0.10
Less than 30 s (at 25 90%response)
Less than 0.1 mg/L
(0 ~ 45) C (automatic)
(0 ~ 45) PPT (automatic)
(80 ~ 105) kPa (automatic)
Be full of waterandAir saturated with airandThe water;
Polarographic
|
Temperature
|
Measuring range
Resolving power
Accuracy
|
-10 ~ 110 DEG C
0.1 C
5~ 60 C Range: + 0.4 + - 0.8 DEG C RestRange
|
Other technical parameters
|
Data storage
Storage content
Power Supply
Communication interface
Sizeandweight
QualityandSecurity authentication
|
600group
Numbering of measurement, measured value, temperaturevalue, ATCorMTCstate, Measurement date, Measurement time
DC9V/300mA
RS232
160 * 190 * 70mm/880g
ISO9001:2000, CEandCMC
|
Working conditions
|
Environmental Sciencetemperature
Environmental Sciencehumidity
IPGrade
|
10 ~ 30 DEG C
Less than or equal to 80%
IP54 Dust prevention and splash prevention
|