The main performance parameters.
Five channel (potassium, sodium, lithium, calcium, barium)
Five kinds of element concentration direct reading
- menu type keyboard
And the automatic calculation of correlation coefficient
And can be equipped with printer
And can be connected to the computer
The size of the flame, pre selected
- with flameout protection device
Response time: less than 8s
- repeat : Less than or equal to 3%
- sample aspiration volume: 6mL/min
- Light: interference filter
Display mode: LED LCD screen
- stability: 15s within = 3%; 5min = 5%
The detection limit. : K: = 0.004mmol/L ; Na: = 0.008mmol/L;
Li: = 0.015mmol/L; Ca: = 0.050mmol/L; Ba: = 0.044mmol/L
- linear error : K: = 0.005mmol/L ; Na: = 0.03mmol/L;
Li: = 0.021mmol/L; Ca = 0.075mmol/L; Ba:: = 0.066mmol/L
- Test Range : K:0 ~ 100ug/ml ~ 160ug/ml; Na:0;
Li:0 ~ 100ug/ml ~ Ba:0 ~ Ca:0; 1000ug/ml; 3000ug/ml
- filter transmittance:
Potassium, lithium, calcium, barium, the absolute value of peak wavelength error Less than 7Nm ; Sodium: absolute value of peak wavelength error Less than 5nm
Potassium, sodium, lithium, calcium, barium, half width Less than 15nm ; ?? Potassium, sodium, lithium, calcium, barium: background transmittance Less than 0.5%;?
?
- Applications
Ceramic, glass and cement - soil and fertilizer
Medical and clinical refractory materials - third party Professional Colleges